SunblockSPF
The language and labelling of sun protection
Edition of 7 August 2026
Testing and standards

What an in vivo SPF test actually involves

A panel of human volunteers, a weighed quantity of product, a solar simulator and a graded visual endpoint. The single integer on a pack is the compressed result of a long procedure.

Standards13 min
In short

A sun protection factor is determined on human volunteers. A weighed quantity of product, two milligrams per square centimetre, is spread over marked areas of skin, a solar simulator delivers a graduated series of ultraviolet doses to protected and unprotected sites, and a trained assessor reads the minimal erythemal dose on each the following day. The individual factors are averaged and the standard requires a stated statistical precision before the result can be reported.

A quartz cuvette in a beam. Every printed figure begins with a controlled source and a measured attenuation.
A quartz cuvette in a beam. Every printed figure begins with a controlled source and a measured attenuation.

Why the test uses people

The sun protection factor is defined against a biological endpoint: the reddening of skin. That definition forces the method. A quantity measured on an instrument would be a different quantity, and the number on a pack would mean something else.

This is a source of persistent tension in the field. Human panel testing is slow, expensive and variable, and there has been sustained work on in vitro alternatives for decades. What has kept the in vivo method in place is that the thing being measured is defined in terms of skin, and any substitute has to be validated against the method it would replace.

The consequence for a reader is that the number on a pack came from people. Somewhere, a panel of volunteers was recruited, screened, marked out with a template and exposed to a graduated series of ultraviolet doses so that a manufacturer could print an integer.

The panel

Volunteers are recruited against criteria set by the standard: skin type within a defined range, no relevant skin condition, no recent ultraviolet exposure on the test area, no medication that alters photosensitivity, and informed consent. The test area is the back, because it is flat, large, easy to shield and rarely exposed.

The standard sets a minimum number of valid individual results before a factor can be reported, and testing continues, up to a maximum, until the statistical requirement is satisfied. The reason for that construction is that people differ. Two volunteers given the same product at the same quantity will not produce the same individual factor, and the reported result is a mean of a distribution rather than a single reading.

Establishing the baseline

Before any product is applied, each volunteer's unprotected response has to be established, because the factor is a ratio and the denominator is personal.

A series of small sub-areas is exposed to a graduated series of ultraviolet doses, and the sites are read after a defined interval, conventionally the following day. The lowest dose that produces a just perceptible reddening across the whole exposed sub-area is that individual's minimal erythemal dose. It is read visually, by a trained assessor, under standardised lighting.

Visual assessment is the part of the procedure that surprises people who assume the whole thing is instrumental. It is deliberate, and it is why assessor training and consistency form part of the method.

Application, which is the controlled variable

The product is weighed. Two milligrams per square centimetre of the marked test area is dispensed, distributed in small dots across the area and then spread with a gloved finger using a specified technique and a specified time, and the film is allowed to settle before exposure.

Every element of that sentence is a control. The quantity is weighed rather than estimated because the factor depends on it steeply. The spreading is timed because a film worked for longer behaves differently. The settling period exists because some formulations change as they dry.

This is the point at which the laboratory and the beach diverge completely, and it is why the application quantity the number assumes is the most consequential condition attached to the figure.

The solar simulator

The source is not the sun. It is a xenon arc solar simulator with filtering that gives it a specified spectral output within defined tolerances across the ultraviolet range, checked by spectroradiometry.

Standardising the source is what makes results comparable between laboratories and between years. Sunlight varies with elevation, season, latitude, altitude and atmospheric conditions, none of which can be held constant. A simulator can be held constant, and the standard specifies how constant.

The trade off is that the simulator's spectrum is a defined approximation of one condition of solar ultraviolet, and the factor is a property measured under that approximation. This is not a criticism. It is the reason the number is reproducible at all.

What is controlled in the test and what replaces it in use
ElementControlled in the test byIn ordinary use
QuantityWeighing to two milligrams per square centimetreNot measured
SpreadingSpecified technique over a specified timeBy hand, variable
SourceSolar simulator with verified spectral outputSunlight, varying continuously
SiteFlat prepared area on the backCurved, mobile, partly covered
IntervalDefined settling period before exposureOften none
ReadingTrained assessor, standardised lightingNot read at all

Framework of this publication. It describes how published rules and guidance fit together and is not a measurement, a survey or a study.

Exposure and reading

Protected sub-areas receive a graduated series of doses, scaled around the expected factor so that the reddening threshold falls within the series. The unprotected sub-areas receive their own series. Everything else on the back is shielded.

Sites are read after the defined interval by the trained assessor, and each volunteer's individual factor is the ratio of the protected threshold dose to the unprotected one. Individual results that fail the validity criteria in the standard are excluded rather than averaged in.

The reported factor is the arithmetic mean of the valid individual results, with a confidence interval calculated around it.

The statistics, and why the pack hides them

A panel mean has an uncertainty, and the standard requires the result to reach a stated precision before it can be reported. Where the precision is not met with the minimum panel, more volunteers are tested until it is, up to a maximum.

The printed figure discards all of this. The tested result is rounded down to the nearest permitted labelling figure, so a product tested somewhat above 30 is labelled 30. The confidence interval is not printed, the panel size is not printed, and the laboratory is not named.

Whether that is the right choice is a genuine question. Printing the interval would communicate the true precision of the measurement and would also be incomprehensible on a pack. The current convention chooses simplicity, and the cost is that a measured quantity with a known uncertainty is presented as a fact with none. That is examined further in measurement uncertainty and what a tested figure is.

The ethical dimension

A test that involves exposing volunteers to ultraviolet doses sufficient to produce visible reddening carries obvious ethical considerations, and they are managed through the same mechanisms as other human studies: screening, informed consent, dose limitation, ethical review and the exclusion of volunteers for whom the exposure would be inappropriate.

It is also the main practical driver behind the long standing effort to develop validated in vitro alternatives. Any replacement has to produce results equivalent to the in vivo method across a wide range of formulation types, which has proved harder than it sounds, because formulations differ in how they form a film on a substrate.

What the method costs, and why that shapes the market

An in vivo determination requires recruitment, screening, a controlled facility, a calibrated simulator, trained assessors and two visits per volunteer. It is not a quick test and it is not a cheap one, and both facts have consequences that reach the shelf.

Every reformulation that could affect the ultraviolet filter system needs retesting. A change of filter, of concentration, of emulsion type or in some cases of pack format is a change to the thing measured, and a figure obtained on the previous version supports a claim about the previous version. That is one reason product ranges are more stable than the pace of cosmetic marketing would suggest.

It also explains why the number is treated as an asset by the businesses that hold it. A determination is expensive enough that its result is defended, and the temptation at the margin is to describe it more generously than the report supports rather than to test again. That temptation is the origin of a large share of the claim problems this publication examines, and it is worth naming rather than pretending that claims go wrong at random.

What this means for reading a label

Three things follow. The figure is real, in the sense that it came from a measurement with a written method. The figure is narrow, in that it describes a property under one set of controlled conditions. And the figure is compressed, in that a distribution with an uncertainty has been reduced to an integer.

A reader who holds all three in mind reads a sun protection pack correctly. A reader who holds only the first treats a laboratory property as a promise about an afternoon.

What to take from this

Volunteers, a weighed film, a standardised source, a visual endpoint and a statistical requirement. The number on the pack is the compressed, rounded down output of that procedure, and everything the procedure controlled for is the list of things that will not be controlled in use.

Questions

Is SPF tested on people or in a laboratory instrument?

On people. The factor is defined against the reddening of skin, so the reference method uses a panel of human volunteers, a weighed application and a solar simulator, with the endpoint read visually by a trained assessor.

How many volunteers are used?

The standard sets a minimum number of valid individual results and requires the panel mean to reach a stated statistical precision. Where it does not, further volunteers are tested up to a maximum specified by the method.

Is the light source real sunlight?

No. It is a solar simulator with a specified spectral output verified by measurement. Sunlight cannot be held constant, and the constancy is what makes results comparable between laboratories.

Why is the endpoint assessed by eye?

Because the factor is defined against visible reddening. Assessor training and standardised viewing conditions are part of the method for that reason.

Why is the confidence interval not on the pack?

Labelling convention reports a single rounded down figure. Printing the interval would communicate the true precision of the measurement and would be very difficult to present usefully in the space available.

Sources

Links go to public institutions and published instruments. They are cited because they are public and checkable, not as endorsement of this publication.

Editorial disclosure

This article contains no commercial links. No brand, product, laboratory, consultancy or supplier has paid for, been offered, or been promised any mention in it, and none is named. Outbound links go to public institutions and published instruments only. Published by Northbank Media.

This publication does not name, rate, rank, review or recommend sunscreen products. It rules on wording, not on packs.

The wording note

One email a fortnight on sun protection claim language: what changed in the codes, what a published ruling turned on, and one phrase examined in full. Written for people who write, approve or check labels and advertising. Unsubscribe whenever you like.

Each issue carries one clearly labelled sponsor line, sold from a published rate card that cannot influence any claim ruling. We use your address for the newsletter and nothing else. See privacy.